Adjusting and Calibrating the Differential Pressure Transmitter using the MT220
Setting 0% and 100% by Applying Actual Pressure (Automatic Setting)

Error Display

Example
Lit during %ERROR
display
Lit when AUTO
0, 100% is set
42
1.
Apply pressure which corresponds to 0%.
2.
Press the Auto %.
3.
Apply pressure which corresponds to 100%.
4.
Press the Auto 100%.
5.
Press the %ERROR key to activate % display mode.
Note
• Any 0% and 100% pressure values can be set as long as they are within the display range
and satisfy the condition 0% value < 100% value.
• If the applied pressure is outside the display range, it can be set as the 0% or 100% value
even if the AUTO 0% or AUTO 100% has been pressed.
• If the 0% value is greater than the 100% value,
%ERROR is pressed. In this case, enter the correct 0% and 100% values.
Automatic setting (using the AUTO 0% and AUTO 100%) is possible only when the
instrument is in normal measurement display mode.
• If the instrument is in relative display mode, % display will not be activated even if the
%ERROR is pressed.
The output voltage (or current) can be checked against the pressure input to
the differential pressure transmitter. The %ERROR value represents the value
obtained by subtracting the input pressure value displayed as a percentage from
the current/voltage value displayed as a percentage. .
The pressure values are shown under % display, and the DMM values are
shown under %ERROR display.
When 20 kPa is set as 0% and 100 kPa as 100%:
When the applied pressure is 40 kPa, 0% (25 - 25) will be displayed as the
%ERROR if the current is 8 mA, and 12.5% (37.5 - 25) will be displayed as
%ERROR if the current is 10 mA.
1.
Make the necessary settings for % display (Refer to the diagram of the previous
page).
2.
Press the %ERROR to activate %ERROR display mode.
Note
• If the %ERROR key is pressed to activate % display mode, the value measured by the DMM
function is also displayed as a percentage. However, this does not apply if the DMM function
is OFF.
• The display switches as follows each time the %ERROR key is pressed.
DMM function ON: normal measurement display→% display→%ERROR display→normal
DMM function OFF: normal measurement display→% display→normal measurement display
Error
1 0 0 %
measurement display
will be displayed when the
kPa
%
V
2 4 V - O L -
mA
%
Lit during an overload
Lit during 24 V output
Mark indicates selected units
IM 767361-01E