Specifications; Pressure Measurement Section (767351/767361/767381/767353/767363/767383/ 767355/767365/767385) - YOKOGAWA MT210 User Manual

Digital manometer
Table of Contents

Advertisement

Specifications

Pressure Measurement Section (767351/767361/767381/767353/767363/767383/
767355/767365/767385)
Model Code
Pressure Type
Measuring Range
(Accuracy guaranteed)
Maximum Display
*1
Accuracy
(Zero CAL)
6 months after calibration
Normal measurement
mode
(Averaging ON)
*2
Medium-speed
measurement mode
(added to normal
measurement mode)
*2
High-speed mode
(added to medium-speed
measurement mode)
One year after calibration ±0.01% of FS
(Represented by adding
the value shown on the
right to the accuracy
percentages given above for
six months after calibration.)
*3
Display Update Interval
Normal measurement
mode
*2
Medium-speed
measurement mode
*2
High-speed
measurement mode
*4
Response Time
Normal measurement
mode
*2
Medium-speed
measurement mode
*2
High-speed
measurement mode
Resolution
Internal Capacity
Input Tolerance
Temperature Coefficient
Effect of Inclination
Weight (Main unit)
*1 Operating conditions: Temperature 23 ± 3°C, humidity 20 to 80% RH, power voltage rating ±5%, positioned horizontally.
Refer to Yokogawa's pressure standard.
*2 Measurement mode can be switched between normal, medium-speed, and high-speed on the 76738
*3 Data output rate via communications is the same as the display update rate.
*4 Measurement conditions of response time:
• Time from the start of change until the value settles within ±1% of the convergence value.
• Release from full scale value to atmospheric pressure. (Input section is no load condition. From 0 to atmospheric pressure
for the absolute pressure model.)
• Measure using D/A output.
68
767351/767361/767381
Gauge pressure
Positive pressure: 0 to 10 kPa
Negative pressure: –10 to 0 kPa Negative pressure: –80 to 0 kPa Negative pressure: –80 to 0 kPa
–12.0000 to 12.0000 kPa
Positive pressure
±(0.01% of rdg +0.15% of FS)
Negative pressure
±(0.2% of rdg +0.1% of FS)
±0.02% of FS
±0.04% of FS
250 ms
100 ms
100 ms
2.5 s or lower
200 ms or lower
200 ms or lower
0.0001 kPa
Approx. 10 cm
3
2.7 kPa abs to 500 kPa gauge
(Up to 50 kPa gauge for the 767381)
Zero drift ±0.0015% of FS/°C
Sensitivity ±0.001% of FS/°C
at 90° inclined forward or
backward
Zero drift ±0.1% of F.S.
30° inclined right or left
Zero point ±2.5% of FS
767351: approx. 8 kg
767361/767381: approx. 8 kg
767353/767363/767383
Gauge pressure
Positive pressure: 0 to 130 kPa Positive pressure: 0 to 700 kPa
to 156.000 kPa
Positive pressure 20 to 130 kPa: Positive pressure:
±(0.01% of rdg +3 digits)
0 to 20 kPa: ±5 digits
Negative pressure
±(0.2% of rdg +0.1% of FS)
±0.02% of FS
±0.03% of FS
±0.005% of FS
250 ms
100 ms
100 ms
2.5 s or lower
200 ms or lower
50 ms or lower
0.001 kPa
Approx. 10 cm
3
2.7 kPa abs to 500 kPa gauge
Zero drift ±0.001% of FS/°C
Sensitivity ±0.001% of FS/°C
at 90° inclined forward or
backward
Zero drift ±0.01% of FS
30° inclined right or left
Zero point ±0.2% of FS
767353: approx. 7 kg
767363/767383: approx. 6.5 kg 767355: approx. 8.5 kg/
767355/767365/767385
Gauge pressure
to 840.00 kPa
±(0.01% of rdg +0.005% of FS)
Negative pressure
±(0.2% of rdg +0.1% of FS)
±0.02% of FS
±0.03% of FS
±0.005% of FS
250 ms
100 ms
100 ms
2.5 s or lower
200 ms or lower
70 ms or lower
0.01 kPa
Approx. 10 cm
3
2.7 kPa abs to 3000 kPa gauge
Zero drift ±0.001% of FS/°C
Sensitivity ±0.001% of FS/°C
at 90° inclined forward or
backward
Zero drift ±0.01% of FS
30° inclined right or left
Zero point ±0.05% of FS
767385: approx. 6.5 kg
767365: approx. 8 kg
only.
IM 767361-01E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mt210fMt220