4.2.2 Description of the error message system
The error message is shown in the "time" display through a two-digit number (19). At the same time
the word "error" (20) is indicated in the display (see photo 20).
20
Photo 20
4.2.3

Error messages

Errors that may be indicated in the LCD display:
Error No.:
Description
01
Rotor Imbalance
02
Imbalance sensor is defective
14
Leap of speed is too big between measurements
30
Motor Error
33
Open lid while the motor is running
34
Lid contact defective
55
Overspeed
60
Undervoltage in the intermediate circuit
70
Blackout of the relay