Installation; Technical Data - Hermle Z 206 A Instruction Manual

Compact centrifuge
Hide thumbs Also See for Z 206 A:
Table of Contents

Advertisement

1.2.1.3

Installation

Follow these steps:
• Check whether power supply corresponds with the one named on the manufacturer's rating
label which is mounted on the rear panel.
• The line voltage circuit breaker is max. 10 A (type K) slow release for commonly used instruments.
• In case of emergency, there must be an emergency switch off installed outside the room in order to
disconnect the power supply of the unit.
(• Remove the transport spacer blocks from the motor shaft (see chapter 2.2.2).)
The outlet for the power cord must be easy to reach and to disconnect!
1.3

Technical Data

Manufacturer
Type / Model
Dimensions
Width
Depth
Height
Weight
Noise level (max.)
Max. speed
Max. volume
Max. RCF
Admissible density
Admissible kinetic energy
Electrical connection AC
Current
Connected load
Interference suppression
Test obligations
To be filled in by purchaser:
Inventory-No.:
Check-No.:
Location:
Maintenance contract:
Your service department
Your agent
Page 6 of 24
HERMLE Labortechnik GmbH
Z 206 A
28 cm
37 cm
26 cm
15 kg
60 +2.0 dB (A)
6000 rpm
6 x 50 ml
4185 x g
3
1.2 kg/dm
1694 Nm
230 V / 50 Hz 1 ph
0.55 A
100 Watt
VDE 0871, Funkentstörgrad B
Labnet International
PO Box 841, Woodbridge, NJ
Ph: 732-417-0700
INSTRUCTION MANUAL Z 206 A
1 PRODUCT DESCRIPTION
120 V / 60 Hz 1 ph
1.1 A
100 Watt
www.bioecoms.com