Pre-Selection Of Speed/Rcf-Value - Hermle Z 206 A Instruction Manual

Compact centrifuge
Hide thumbs Also See for Z 206 A:
Table of Contents

Advertisement

2.2.4 Pre-selection of speed / RCF-value
By pressing the key "rpm/rcf" (6) the pre-selection of the speed is activated. By pressing the key
once the word "rpm" (12) flashes.
By pressing the key once again the pre-selection of the centrifugal forces may be chosen. Flashing
the word "rcf" (13).
You can set the desired values with the adjusting knob (7). In the display (14) the regulated value
is shown permanently, before, during and after the run.
The speed is adjustable between 200 rpm and maximum revolution of the centrifuge resp. the
maximum permissible revolution of the pre-selected rotor.
It is the same with the pre-selection of the RCF-value. The setting range is between 20 xg and
the maximum permissible centrifugal force of the rotor.
The maximum speed of the Z 206 A is 6000 rpm resp. 4180 xg.
14
12
Photo 13
Maximum speed of the rotors available for the Z 206 A
Attention:
Please also check the maximum permissible speed of your test tubes.
Page 16 of 24
6
Rotor-Number
Max.
Revolution
220.96 V01
6000 rpm
220.97 V01
6000 rpm
220.95 V06
6000 rpm
220.68 V04
3500 rpm
INSTRUCTION MANUAL Z 206 A
13
14
7
RCF
Value
4180 xg
3820 xg
2930 xg
1450 xg
www.bioecoms.com