Diagnostics - ABB Endura APA592 Operating Instruction

Ph/redox orp transmitter
Table of Contents

Advertisement

Endura APA592
pH / Redox (ORP) transmitter

7.3.6 Diagnostics

Menu
Diagnostics
123
Exit
Diagnostics Parameters
Parameter
PV cal alarm lim
Slope alarm lim high
Slope alarm lim lo
Offset alarm Lim +/–
Loop Test
Diagnosis Status
68
Used to perform a loop test (output verification) and indicates
6
the diagnostic status (actual / simulated).
Select
Description
100 % to 150 %. Default = 110 %.
40 % to 100 %. Default = 60 %.
0 to ±10000 mV. Default = ±180 mV.
The current output of the transmitter can be adjusted between 3.8 and
21.5 mA.
This feature is useful to verify that the transmitter output matches the current
in the control room.
Enabled when HART Poll Addr = 0 – see page 69.
Used to determine if diagnostics simulation is active.
When display reads Actual, real-time diagnostics from the transmitter are
active.