Dc Power Supply Connections - ABB Endura APA592 Operating Instruction

Ph/redox orp transmitter
Table of Contents

Advertisement

Endura APA592
pH / Redox (ORP) transmitter

4.2 DC Power Supply Connections

3
Key:
1
Positive +
2
Negative –
3
Earth (Ground)
Fig. 4.2 Power Supply Connections at Transmitter Terminals
Note. HART Communication
Communication with the transmitter is supported by the HART protocol. The communication signal is
modulated on both wires for the power supply line and decoded. The electrical connection is provided at
the (+) and (–) terminals of the transmitter or by the power supply cable installed on-site.
To connect the HART power supply:
1. Route the power supply wiring through one of the three threaded connections in the transmitter
housing.
2. Strip the wires and tin or attach wire end sleeves.
3. Loosen the power supply terminal captive screws (
(3.5 / 4 mm or Size-1).
4. Connect the (+) wire to the (+) terminal
5. Connect the (–) wire to the (–) terminal
6. Connect the cable shield or earth (ground) wire to the earth (ground) terminal
28
GND
+
Connection Label*
(inside End Cap)
1
2
*DC Power Supply Requirements:
Non-Intrinsically safe transmitters:
12 to 42 V DC, 4 to 20 mA
Intrinsically safe transmitters:
12 to 30 V DC, 4 to 20 mA
1
2
,
and
1
.
2
.
4 Electrical Installation
4
5
6
3
7
10
11
2
8
SUPPLY
4-20 mA
+
1
9
APA592-PH
Transmitter Type
Identified Here
(APA592–PH)
3
) using a flat-bladed screwdriver
3
(if required).
OI/APA592–EN Rev. E