Endura APA592
pH / Redox (ORP) transmitter
Step
Procedure
At the calibration results screen(s) check
the new values. Press the
the screen:
or
if a calibration failure message is
7
displayed, reset the buffer values (see
steps 2 to 6).
If the Hold Output screen was selected at
step 1, an option to release the output is
provided at the Release Output screen.
8
Press the
Output state or the
Hold Output state.
Table 7.3 2-Point Manual Calibration (Continued)
60
key to exit
key to retain the Hold
key release the
7 Operator Pages and Menus
Screen
OI/APA592–EN Rev. E