Hazardous Area Relevant Information; Apa592-Ph.a1 - ABB Endura APA592 Operating Instruction

Ph/redox orp transmitter
Table of Contents

Advertisement

Endura APA592
pH / Redox (ORP) transmitter

2.5 Hazardous Area Relevant Information

Note. The explosion-proof or ignition-proof designation is displayed on the agency certification label.
2.5.1 APA592-pH.A1... (Intrinsic Safety)
ATEX/IECEx approved for use in zone 0/20.
Ex ia (Zone 0):
LCIE 11 ATEX 3058 X
IECEx LCI 11.0050X
Designation
Device design: II 1 G Ex ia IIC T4
Housing design: II 1 D Ex iaD A20 IP66 T135 ºC, –20 ºC  T
Safety-relevant data
Input Parameters
Maximum voltage
Maximum input current
Maximum power
Internal inductance
Internal capacitance
Table 2.1 Intrinsic Safety Input Parameters
Output Parameters
Open-circuit voltage (maximum)
Short-circuit current (maximum)
Maximum output power
Allowed inductance (total)
Allowed capacitance (total)
Table 2.2 Intrinsic Safety Output Parameters
Notes.
Parameters apply to entire system inclusive of cables.
Each specified electrical parameter must be applied individually and in combination. Do not
exceed the maximum values when applying the electrical parameters individually or in
combination.
Both Intrinsic safety and dust ratings are combined when APA592-pH.A1... is ordered.
12
2 Use in Areas Requiring Ignition Protection
 60 ºC
amb
U
= 30 V
i
Ii = 160 mA
P
= 0,8 W
i
L
= 0,5 mH
i
C
= 5 nF
i
U
= 11.8 V
o
I
= 5 mA
o
P
= 15 mW
o
L
= 1 H
a
C
= 1.45 µF
a