Endura APA592
pH / Redox (ORP) transmitter

7.3 Parameters

7.3.1 Easy Setup

Menu
Easy Setup
Exit
Easy Setup Parameters
Parameter
Analyzer Type
pH Sensor Type
Isopotential Pt.
Asymmetric Pt.
Valence
Magnitudes
End Mag
End mV
PV Units
Temp. unit
54
Users with Advanced level access have read / write privileges
1
to the parameters listed.
Users with Standard and Read Only access have read-only
privileges to parameters at Easy Setup level.
Select
Description
Selects the measurement type:
pH
ORP
pION
Ion Conc
Selects the sensor type used for measurement:
Glass
Antimony
Custom
(Isopotential Point)
Adjustable range –20 to 20 pH. Default = 7 pH.
Enabled when pH Sensor Type = Custom.
(Asymmetric Potential)
Adjustable range –2000 to 2000 mV. Default = 0 mV
Enabled when pH Sensor Type = Custom.
Ion valence selection: -3, -2, -1, 1, 2, or 3.
The valence determines the millivolt change per decade of concentration.
The number of magnitudes ranging from 1 to 3 that defines the ion
concentration output.
The ion concentration state functions by associating an end millivolt value to
an end magnitude value.
The end magnitude value can be set to 10, 100 or 1000.
The end millivolt value associated with the end magnitude value.
Sets the process value units (%, ppm, µg/l, ppb, mg/l).
(Temperature Units)
Sets the temperature unit (°C or °F) applied to all transmitter temperature
parameters and values.
7 Operator Pages and Menus
OI/APA592–EN Rev. E
Tx
Type
All
pH
pH
PH
Ion
Conc
Ion
Conc
Ion
Conc
Ion
Conc
Ion
Conc
All