Endura APA592
pH / Redox (ORP) transmitter
4.4 Sensor Cable Color-Coded Connections
Note. ORP (Redox) and Antimony pH sensors do not have temperature compensation and therefore do
not have temperature sensors or related wiring.
4.4.1 Standard Sensors with No Diagnostic Functions
Note. Turn Off diagnostics when connecting these sensors
Connection Information
Transmitter
Terminal
4
Glass / Metal Sensing Electrode
5
Guard
6
Reference
7
Solution Ground
Temperature Compensator
8
(If included with sensor)
Temperature Compensator
9
(If included with sensor)
10
Shield /Shield (if fitted)
Temperature Compensator
11
(If included with sensor)
*Cut back the 7650 / 60 sensor's redundant green wire.
Table 4.2 Standard Sensors with No Diagnostic Functions
30
Function
4 Electrical Installation
pH Sensor Cable Color
2867
AP100
AP300
Clear
Clear
Blue
–
–
–
Black
Black
Black
–
–
–
–
Red
Red
–
White
White
–
–
–
–
Red
Grey
7650/
TB5
60*
Clear
Blue
–
–
Black
Black
–
–
Red
Red
White
White
Yellow
–
Red
–
OI/APA592–EN Rev. E