ABB Endura APA592 Operating Instruction page 66

Ph/redox orp transmitter
Table of Contents

Advertisement

Endura APA592
pH / Redox (ORP) transmitter
Device Setup Parameters
Parameter
Temp. Sensor
Temp. unit
Temp. Comp. Type
Manual Temp
pH Soln. Coeff.
mV Soln. Coeff.
Temp. Sensor Type
Run TC Recognition
TC Recognition Stat
64
Description
Temperature sensor parameters.
Sets the temperature unit (°C or °F) applied to all transmitter temperature
parameters and values.
(Temperature Compensation Type)
Manual
The value in this parameter is used for the referenced temperature
compensation during manual temperature compensation mode
(default value 25 °C (77 °F). Manual is the default parameter if a valid
temperature sensor is not recognized at power-up or when a Run TC
Recognition routine is run (see below).
Auto
Enabled when a temperature sensor is provided with the probe (the
calculation is based on the Nernst equation – see Appendix A.3,
page 98).
Auto is the default parameter if a valid temperature sensor is
recognized at power-up or when a Run TC Recognition routine is run.
Auto Solution
Auto Solution temperature compensation can be selected when a
temperature sensor is connected and is recognized by the device.
This selection provides the ability to include a solution coefficient in
the Nernst equation – see Appendix A.3, page 98.
The value in this parameter is used for temperature compensation during
Manual temperature compensation mode.
Adjustable range –20 to 150 °C (–4 to 302 °F). Default = 25 °C (77 °F).
Enabled when Temp Comp Type = Manual.
Adjustable range –10 to 10. Default = 0 pH / 10 °C (50 °F).
Enabled when Temp. Comp Type = Auto Solution.
Adjustable range –20 to 20. Default = 0 mV / 10 °C (50 °F).
Enabled when Temp. Comp Type = Auto Solution.
Indicates the type of integral temperature sensor detected:
Pt100
Pt1000
Balco 3k
Starts the auto-recognition process for the sensor.
Acceptable types for this process are: Pt100 2-wire, Pt100 3-wire, Pt1000
2-wire, Pt1000 3-wire, Balco 2-wire, Balco 3-wire.
Indicates if the temperature sensor was recognized / not recognized, either
during startup or after execution of the Run TC Recognition function, above.
7 Operator Pages and Menus
OI/APA592–EN Rev. E
Tx
Type
All
pH
ORP /
pION,
Ion
Conc