Endura APA592
pH / Redox (ORP) transmitter
Output Parameters
Open-circuit voltage (maximum)
Short-circuit current (maximum)
Maximum output power
Allowed inductance (total)
Allowed capacitance (total)
Table 2.9 Non-incendive Field Wiring – FM and CSA: Output Parameters
Notes.
Parameters apply to entire system inclusive of cables.
Each specified electrical parameter must be applied individually and in combination. Do not
exceed the maximum values when applying the electrical parameters individually or in
combination.
16
2 Use in Areas Requiring Ignition Protection
U
= 11.8 V
o
I
= 5 mA
o
P
= 15 mW
o
L
= 1 H
a
C
= 1.45 µF
a