Available Rotors - Beckman Coulter Allegra X-30 Series Instructions For Use Manual

Rx only
Table of Contents

Advertisement

Description

Available Rotors

Electrical supply
Maximum heat dissipation into room under
steady-state conditions
Noise level 0.91 m (3 ft) in front of
instrument (approx.)
Installation (overvoltage) category
Pollution degree
a. Temperature range depends on rotor in use and speed (see applicable rotor manual).
b. Normally only nonconductive pollution occurs; occasionally, however, a temporary conductivity caused by condensation must be expected.
Available Rotors
See the applicable rotor manual for information on rotor use, care and maintenance, and rotor
accessories.
Rotor Profile
Description
F2402H
Fixed Angle
45° Angle
r
= 82 mm
max
FX301.5
Fixed Angle
45° Angle
r
= 100 mm
max
F0630
Fixed Angle
30° Angle
r
= 78 mm
max
F0850
Fixed Angle
25° Angle
r
= 94 mm
max
F0685
Fixed Angle
25° Angle
r
= 97 mm
max
1-10
Class I
3311 Btu/h (0.97 kW)
£68 dBa
II
b
2
Refrigerated
Max
Max RCF
RPM
( g)
18,000
29,756
a
b
16,000
28,672
18,000
28,305
11,400
13,682
10,000
10,864
Nonrefrigerated
Max
Max
RPM
RCF
( g)
16,000
23,511
13,200
19,515
b
16,000
22,364
9000
8528
8000
6953
6  85
Max
Rotor Part
Capacity
Number/
(mL)
Rotor Manual
Number
24  1.5/2.0
361171
GS-TB-021
30  1.5/2.0
392274
MMR-TB-002
6  30
361231
GS-TB-014
8  50
346640
GS-TB-003
364650
GS-TB-008
PN B05529AE

Hide quick links:

Advertisement

Table of Contents
loading