Procedure - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

1
HP 53181A
Counter
Channel 2
Figure 1-15. Option 015/030/050/124 Channel 2 Frequency Accuracy Test Setup
2
Chapter 1 Performance Tests
Test 5: Option 015/030/050/124 Channel 2 Frequency Accuracy (HP
53181A Only)

Procedure

Connect equipment as shown in Figure 1-15.
50Ω
Feedthrough
Ref. In
(rear panel)
Set the HP 8663A Synthesized Signal Generator to output
100 MHz at −17.0 dBm.
Verify the Counter reads 100 MHz ±70 mHz.
1-64
HP 8340B
Sweep Oscillator
HP 8663A
Synthesized Signal Generator
Assembly-Level Service Guide
Frequency Standard
(switch set to EXT)
Output
Input
(rear panel)
Output
N-to-BNC
Connector