Volt Peak - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)

Volt Peak 2

1
Remove the signal from Channel 1 of the Counter and connect it
to Channel 2.
2
Press Other Meas key until VOLT PEAK 2 is displayed.
The Counter should display -1.00 ±0.12V for the negative peak of the sine
wave. This reading is displayed on the left side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 2a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the
sine wave. This reading is displayed on the right side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 2b).
3
Disconnect the test setup.
Assembly-Level Service Guide
1-39