Chapter 2 Service
Troubleshooting the Counter
Table 2-4 lists A4 power supply voltages, the corresponding wire color,
and the tolerances.
Table 2-4. A4 AC Power Supply Voltages
Wire Color
Red
Red
Black
Black
Yellow
Black
Blue
5
After verifying the correct voltage and that all calibrations
passed, proceed to the following section.

Overview of the Self-Test Routines

The Utility menu includes the following three self-test menu items:
TEST LOOP:
TST PRINT:
TEST:
The Counter includes several self-tests. Some self tests are performed at
the power-up of the Counter. Some of these self tests can also be invoked
from the Utility menu. Other tests can also be invoked from the Utility
menu, but a slightly different test is performed. For example, a power-up
self test might work regardless of connections to the inputs while a Utility
menu self test may require a specific signal to be connected. Some self tests
can only be invoked from the Utility menu, for example, the keypad test.
You can specify that the tests should loop forever ( TEST LOOP: ON or
OFF ) or until any key is pressed.
Assembly-Level Service Guide
Voltage
Tolerance
+5.0 V
±0.250 V
+5.0 V
±0.250 V
0.0 V
______
0.0 V
______
+12.0 V
±0.6 V
0.0 V
______
-12.0 V
±1.2 V
2-47