NOTE
Chapter 2 Service
About the HP 53131A/132A Calibration Menu
The Calibration menu, illustrated on page 2-8, allows you to:
view the calibration security status of the Counter (CAL SECURE or
CAL UNSECURE)
initiate the Counter's calibration routines, which can perform
automatic calibration of the voltage offset and gain of the front-end
input amplifiers (CAL: OFFS 1 / 2 ?, GAIN 1 / 2?), of the optional
timebase (CAL: TIMEBAS?), and of the differences in electrical
path-length between channels 1 and 2 (CAL: TI QUIK? and
CAL: TI FINE?)
secure against calibration by entering the security code while the
Counter is unsecured (CODE: )
unsecure for calibration by entering the security code while the
Counter is secured (CODE: )
change the security code by entering a new code while the Counter is
unsecured (CODE: )
view the calibration count, which indicates the number of times that
the Counter has been calibrated (CAL COUNT?). The calibration
count increments up to a maximum of 32,767 after which it wraps
around to 1. A calibration count of 0 indicates that the Counter is
completely uncalibrated. (The calibration count is unaffected by
interpolator calibration.)
get Help (HELP: ) with any of the following:
how to calibrate (CAL?),
how to secure (SECURE?), and
how to change the code (CODE?)
A good precaution is to use an HP-IB program (see the programming
example titled "To Read and Store Calibration Data" in Chapter 3 of the
Programming Guide) to read and store the calibration factors prior to
initiating any calibration.
Assembly-Level Service Guide
2-7