The Hp 53181A Calibration Menu Tree - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

NOTE
Chapter 2 Service
About the HP 53181A Calibration Menu

The HP 53181A Calibration Menu Tree

Turn power off, press and hold Scale & Offset key, then press POWER
key to access this menu.
3
This menu item appears and calibration is permitted only if calibration is unsecured. Enter in the correct code
to change calibration to secure.
4
Timebase can be automatically calibrated only if the timebase option is installed.
Assembly-Level Service Guide
POWER
Scale &
Offset
On / Stby
CAL SECURE
CAL UNSECURE
CAL: OFFS1?
3
CAL:
GAIN1?
CAL:
TIMEBAS?
4
CODE: 0
CAL COUNT?
HELP: CAL?
HELP:
SECURE?
HELP:
CODE?
2-27