To Calibrate The Gain For Channels 1 And 2 - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 2 Service
The HP 53131A/132A Calibration Procedures

To Calibrate the Gain for Channels 1 and 2

1
Press any one of the arrow keys until CAL: GAIN 1 ? is displayed.
2
Press Enter key.
A scrolling message is displayed. Follow the instruction by performing the
following steps.
3
Connect a BNC tee connector to Channel 1 of the Counter.
4
Connect the Counter, dc power supply, and digital multimeter as
shown in Figure 2-2.
DC Power Supply
-
-
+
Figure 2-2. Gain Calibration Setup
5
Adjust the power supply until the digital multimeter displays
+ 5.000 V ± 0.001 V.
6
Disconnect the digital multimeter from the Counter to prevent
noise from being introduced into the measurement.
7
Press Enter key.
The Counter momentarily displays CALIBRATING , and then it should
display GAIN 1 PASS .
If the fail message is displayed, refer to the troubleshooting section in this
chapter.
2-18
HP 53131A/132A
Counter
Digital Multimeter
BNC Tee
Assembly-Level Service Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53181a53131a

Table of Contents