Chapter 1 Performance Tests
HP 53131A/132A Performance Test Record (Page 3 of 6)
HP 53131A/132A Performance Test Record
(Page 3 of 6)
Test
Number
Test Description
4
Channels 1 and 2
Frequency
Accuracy
(HP 53131A/132A
Only)
1 MHz Test
(HP 53131A):
Channel 1
Channel 2
1 MHz Test
(HP 53132A):
Channel 1
Channel 2
4
100 MHz Test
(HP 53131A):
Channel 1
Channel 2
100 MHz Test
(HP 53132A):
Channel 1
Channel 2
Assembly-Level Service Guide
Complete Performance Tests (Continued)
Minimum
999,999.999 Hz
999,999.999 Hz
999,999.9998 Hz
999,999.9998 Hz
99.999,999,93 MHz
99.999,999,93 MHz
99.999,999,980 MHz
99.999,999,980 MHz
Actual
Reading
1. __________
1,000,000.001 Hz
2. __________
1,000,000.001 Hz
3. __________
1,000,000.0002 Hz
4. __________
1,000,000.0002 Hz
5. __________
100.000,000,07 MHz
6. __________
100.000,000,07 MHz
7. __________
100.000,000,020 MHz
8. __________
100.000,000,020 MHz
Maximum
1-43