Chapter 1 Performance Tests

Test 1: Trigger Level (HP 53181A Only)

Test 1: Trigger Level (HP 53181A Only)
This test verifies the Trigger Level accuracy of the HP 53181A Frequency
Counter.
Equipment
HP 3325B Synthesizer
HP 10503A BNC Cable
Counter Setup
1
Cycle the POWER key to preset the Counter.
2
Set CHANNEL 1 input condition to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000V
50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off