Chapter 2 Service
The HP 53131A/132A Calibration Procedures
5
Press Enter key.
The Counter displays CALIBRATING , and then it should display
TI CAL PASS .
If the fail message is displayed, refer to the troubleshooting section in this
chapter.
6
Disconnect the calibration setup.
This completes the TI QUIK? calibration procedures.
Depending on which timebase the Counter contains, proceed to the
procedure "To Calibrate the Standard Timebase" (page 2-23) or the
"To Calibrate the High Stability Timebase Option (Medium, High, or
Ultra-High)" procedure (page 2-24) to calibrate the timebase.
CAL: TI FINE? Calibration
1
Connect the HP 8130A Pulse Generator, HP 59992A J06 Time
Interval Calibrator, and HP 53131A/123A Counter as shown in
Figure 2-4.
Equipment
HP 8130A Pulse Generator (or equivalent)
HP 59992A J06 Time Interval Calibrator (or equivalent)
HP 10503A BNC Cable (3)
HP 8130A
Pulse Generator
Output
Figure 2-4. Calibration Setup for CAL: TI FINE?
Assembly-Level Service Guide
In A
B
1
2
3
4
HP 59992A J06
Time Interval Calibrator
HP 53131A/132A
Counter
Stop
Start
2-21