HP 53131A/132A
Counter
Channel 1
Figure 1-10. Peak Volts Test Setup
2
Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
Volt Peak 1
Connect the equipment as shown in Figure 1-10.
Channel 2
Set the HP 3325B to output a 2 MHz, 2 Vp-p sine wave.
The Counter should display −1.00 ±0.12V for the negative peak of the sine
wave. This reading is displayed on the left side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 1a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the
sine wave. This reading is displayed on the right side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 1b).
1-38
HP 3325B
Synthesizer
Assembly-Level Service Guide