Test 4: Option 015/030/050/124 Channel 2 Frequency Sensitivity

2.8 to 3.0 GHz Sensitivity for Option 030 Only

Disconnect the HP 8663A Synthesized Signal Generator from the
Counter (see Figure 1-14), and connect the HP 8340B Sweep
Oscillator to Channel 3 of the Counter.
Change the settings of the Sweep Oscillator to 3.0 GHz, −40 dBm.
Increase the power level in 1 dB steps until the Counter displays
3.0 GHz.
The Counter should display 3.0 GHz at ≤ −21.0 dBm.
Sweep the frequencies from 3.0 GHz to 2.8 GHz.
The Counter should read frequencies from 3.0 to 2.8 GHz at ≤ −21.0 dBm.
Record the actual reading in the Performance Test Record
(Test 4, Line 3).
Disconnect the test setup.
200 MHz to 5.0 GHz Sensitivity for Option 050 only
Connect the 8340B Sweep Oscillator to Channel 2 of the counter.
Change the settings of the Sweep Oscillator to 5.0 GHz, −30 dBm.
Increase the power level in 1 dB steps until the Counter displays a
stable reading of approximately 5.0 GHz.
The Counter should display 5.0 GHz at ≤ −23 dBm.
Use a power meter and sensor to ensure the accuracy of the power being
input.
Sweep the frequencies from 200 MHz to 5.0 GHz.
The Counter should read frequencies from 200 MHz to 5.0 GHz at ≤ −23 dBm.
Record the actual reading in the Performance Test Record
(Test 4, Line 4).