Procedure - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Chapter 4 Retrofitting Options
To Retrofit A6 High Stability Timebase Assembly (Options 001, 010, and
012)

Procedure

1
Turn off the Counter and remove power cord.
2
Remove the cover.
See the section titled "To Remove the Cover" in Chapter 3 of this guide.
3
If you are retrofitting the Option 001 Medium Stability Timebase
Assembly, go to step 5.
If you are retrofitting the Option 010 High Stability or Option 012
Ultra-High Stability assembly, go to step 4 (the next step).
4
Plug the oven oscillator into the socket on A6 Timebase Support
Board, and secure the oven oscillator to A6 using the two Pozidriv
screws (H7) as shown in Figure 4-2.
5
Plug the oven oscillator into the socket on A6 Timebase Support
Board, and secure the oven oscillator to A6 using the two Pozidriv
screws (H7) as shown in Figure 4-2.
6
If you are retrofitting the Option 001 Medium Stability Timebase,
remove the jumper cap on P1 of A6 Timebase Support Board as
shown in Figure 4-2.
If you are retrofitting the Option 010 High Stability and Option
012 Ultra-High Stability timebases, DO NOT remove the jumper
cap on P1.
7
Mount A6 High Stability Timebase Assembly on the left side using
three TORX 10 screws (H1) as shown in Figure 4-2.
8
Plug the power supply assembly's two-wire cable into the
three-prong connector (J2) on the high stability timebase
assembly as shown in Figure 4-2 (if the two-wire cable exists).
9
Connect A6 High Stability Timebase Assembly cable to connector
J9 on A1 Motherboard Assembly as shown in Figure 4-2.