Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
8
Repeat steps 1 through 4 for Channel 1.
9
Record the actual reading in the Performance Test Record
(Test 3, Line 4).

200 MHz to 225 MHz Sensitivity for HP 53131A/132A

1
Leave the signal generator connected to Channel 1. Change the
signal generator settings to 225 MHz at −30 dBm.
2
On the Counter, press Run key.
3
Increase the power level by 0.1 dB steps until the Counter
displays a stable reading of 225 MHz.
The Counter should read 225 MHz when the input signal level is
≤ −15.0 dBm (40 mVrms).
4
Sweep the signal generator frequency from 225 MHz to 200 MHz.
The Counter should read frequencies from 225 MHz to 200 MHz when the
input signal level is ≤ −15.0 dBm (40 mVrms).
5
Record the actual reading in the Performance Test Record
(Test 3, Line 5).
6
Connect the signal generator to Channel 2.
7
On the Counter, press Freq & Ratio key until Frequency 2 is
displayed.
8
Repeat steps 1 through 4 for Channel 2.
9
Record the actual reading in the Performance Test Record
(Test 3, Line 6).
10
Disconnect the test setup.
1-26