HP 53132A Assembly And Service Manual page 245

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

General Information
Save and Recall:
Rack Dimensions (HxWxD):
Weight:
Power Supply Voltage:
Power Requirements:
Operating Environment:
Storage Environment:
Remote Interface:
HP-IB Interface Capabilities:
Remote Programming Language: SCPI-1992.0 (Standard Commands for Programmable Instruments)
Safety:
EMC:
RS-232C:
Note on Pin 4:
1 2 3 4 5
6 7 8 9
Chapter 8 HP 53181A Specifications
General Information
Up to 20 complete instrument setups may be saved and recalled later. These setups are retained when
power is removed from the counter.
88.5 mm x 212.6 mm x 348.3 mm
3.5 kg maximum
AC Line Supply
100 to 120 VAC ±10% - 50, 60 or 400 Hz ±10%
220 to 240 VAC ±10% - 50 or 60 Hz ±10%
AC Line Voltage Selection: Automatic
170 VA maximum (30 W typical)
0° C to 55° C
–40° C to 71° C
HP-IB (IEEE 488.1-1987, IEEE 488.2-1987)
SH1, AH1, T5, TE0, L4, LE0, SR1, RL1, PP0, DC1, DT1, C0, E2
Designed in compliance with IEC 1010-1, UL 3111-1 (draft), CAN/CSA 1010.1
CISPR-11, EN50082-1, IEC 801-2, -3, -4
Electrostatic Discharge and Fast Transient/Burst Immunity Testing: When the product is operated at
maximum sensitivity (20 mVrms) and tested with 8kV AD according to IEC801-2 or with 1kV power line
transients according to IEC 801-4, frequency miscounts may occur that will affect measurement data
made during these disturbances.
Radiated Immunity Testing: When the product is operated at maximum sensitivity (20 mVrms) and tested
at 3 V/m according to IEC 801-3, external 100 to 200 MHz electric fields may cause frequency miscounts.
The rear-panel RS-232 connector is a 9-pin connector (DB-9, male). You can connect the universal
counter to any terminal or printer with a properly configured DTE connector (DB-25). You can use a
standard interface cable (HP part number 24542G or 24542H). Data is "output only"; the instrument can
not be programmed via the RS-232 interface.
May be used as either a DTR signal or an indication of measurement in-limit as configured by the Utility menu.
When used as an in-limit indicator, the signal will be high for every measurement within the user set limits.
Pin Number
2
3
4
5
6
*
Assembly-Level Service Guide
DC Supply (Option 002 Only)
10 to 32 VDC, 3-pin male XLR connector
Option 002 may not be ordered with Option 060
4A initial inrush at 10 VDC
3A max, once stabilized
Type
Input
Receive Data (RxD) (for Xon/Xoff only)
Output
Transmit Data (TxD)
Output
Data Terminal Ready (DTR)
Measurement In-Limit Signal
Signal Ground
Input
Data Set Ready (DSR)
All other pins: no connection
Description
8-11