HP 53132A Assembly And Service Manual page 239

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Measurement Specifications
Frequency, Period
Channel 1 Range:
0.1 Hz to 225 MHz
Channel 2 Range:
Option 015
100 Mhz to 1.5 Ghz
Option 030
100 MHz to 3 GHz
Option 050
200 MHz to 5 GHz
Option 124
200 MHz to 12.4 GHz
(Period 2 selectable only via the HP-IB interface)
For Automatic or External Arming:
(and signals < 100 Hz using Timed Arming)
LSD Displayed:
Frequency
t
res
×
--------------------------- -
or
Gate Time
Period
RMS Resolution:
2
2
(
×
t
2 Trigger Error
+
res
--------------------------------------------------------------------- -
Gate Time
5
t
: 650 ps typical
res
For Automatic Arming:
Gate Time
where N = 1 for Ch1 Frequency < 1 MHz
4 for Ch1 Frequency > 1 MHz
128 for Ch2
±
Systematic Uncertainty:
Time Base Error
t
: 350 ps typical
acc
1.25 ns worst case
Trigger: Default setting is Auto Trigger at 50%
The following graphs may also be used to compute errors for Period Measurements. To find the Period error (∆P), calculate the frequency of the
1
input signal (
) and find the frequency error (∆F) from the chart. Then, calculate the period error as:
F
--- -
=
P
.
Chapter 8 HP 53181A Specifications
Measurement Specifications
4.44 ns to 10 s
0.67 ns to 10 ns
0.33 ns to 10 ns
0.2 ns to 5 ns
80 ps to 5 ns
Frequency
)
×
or
Period
N
--------------------------- -
=
Frequency
Frequency
t
±
a cc
×
----------------------
or
Gate Time
Period
Assembly-Level Service Guide
For Time or Digits Arming:
LSD Displayed:
×
2 2 t
re s
-------------------------------------------------------------------------------------------
×
Gate Time
Number or Samples
RMS Resolution:
2
×
(
×
4
t
+
2 Trigger Error
res
-------------------------------------------------------------------------------------------
×
Gate Time
Number or Samples
5
t
: 500 ps typical
res
5
t
: 50 ps typical
jitter
Number of Samples =
Gate Time x Frequency
Gate Time x 200,000
±
Systematic Uncertainty:
Time Base Error
t
: 100 ps typical
acc
300 ps worst case
Trigger: Default setting is Auto Trigger at 50%
Frequency
t
jitter
×
--------------------------- -
+
or
Gate Time
Period
2
Frequency
)
t
jitter
×
--------------------------- -
or
+
Gate Time
Period
(Frequency < 200 kHz)
(Frequency > 200 kHz)
Frequency
t
±
a cc
×
----------------------
or
Gate Time
Period
F
×
P
------ -
P
=
F
8-5