Equipment Required - HP 53132A Assembly And Service Manual

225 mhz counter
Hide thumbs Also See for 53132A:
Table of Contents

Advertisement

Table 1-1. Recommended Test Equipment
Instrument
Synthesizer
Synthesized Signal Generator 100 kHz to 2560 MHz
Pulse Generator
Sweep Oscillator
Primary Frequency Standard
(10 MHz House Standard)
DC Power Supply
Digital Voltmeter
Time Interval Calibrator
Type N-to-BNC Connector
BNC-to-Dual Banana Plug
50Ω Feedthrough (3)
50Ω Coaxial Cable with
BNC connectors (5)
Coaxial BNC Tee Connector
P = Performance Tests
OV = Operational Verification
C = Calibration
T = Troubleshooting
Chapter 1 Performance Tests

Equipment Required

Equipment Required
Required Characteristics
DC to 13 MHz
5 ns pulse width
3.0 GHz
Absolute accuracy >1 x 10
Adjustable to 5.000V
Microvolt accuracy
Effective edge repeatability:
<10 ps
BNC(m) to banana plug
BNC(m) to BNC(m), 48 inches
Some of this equipment may no longer be available as new. Use an equivalent model number.
1-4
Recommended Model
HP 3325B
HP 8663A
HP 8130A or equivalent
HP 8340B
-10
HP 5071A
HP 6234A
HP 3458A
HP 59992A J06
HP 1250-0780
HP 11001-60001
HP 10100C
HP 10503A
1250-0781
Assembly-Level Service Guide
Use
OV, P, T
OV, P, T
OV, C
P, T
C
C
OV, C
C
OV, P
OV, P, C T
OV, P, C
OV, P, C, T
P, C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

53181a53131a

Table of Contents