Adjustment Procedure; Adjustments (Continued); Procedure; Input Amplifier Adjustment Waveform - HP 5384A Operating And Service Manual

Frequency counters
Table of Contents

Advertisement

HP 1740A TRIGGERED
SWEEP OSCILLOSCOPE
HP 5384A and HP 5385A
Adjustments
(I>
00
(I>
@
@
~
gg
©
@(I)
0 0 @
. . _ _ _ _ _
~
0
fi\ :
~
.-
0
gJ~
0
r~·
oiJD'o
~'J
s
@
g
~@
00
@v:v
0
0
®
G
oooo
0
0
@
®
REMOVE HP 5384A/HP 5385A TOP COVER; CONNECT
OSCILLOSCOPE PROBE AT THE JUNCTION OF R19
AND R65E.
RF OUTPUT
EXT 10 MHz
REFERENCE
HP 5384A/HP 5385A
FREQUENCY
COUNTER
BNC CABLE
INPUT A
Figure 5-2. Channel A Input Amplifier Adjustment Setup
0.05V
T
-ov
__ .. .,..,.,
!---o.o2
J.LSec
Figure 5-3. Input Amplifier Adjustment Waveform
5-15.
HP 5384A Channel 8 Input Amplifier
Adjustment Procedure
5-16.
To perform the sensitivity adjustment proceed
as follows:
a.
Connect the HP 8656A Signal Generator as
shown in
Figure 5-4.
b.
Cycle the power switch from STBY to ON and
press the FREQ B button on the HP 5384A front panel.
c. Set the HP 8656A to output a 200 MHz sinewave
at 6 mV rms. Adjust A 1 R32 for a stable display count.
Vary the frequency from 200 MHz to 50 MHz and
verify that the counter shows a stable display.
d.
This completes the adjustment for Channels A
and B; switch the counter to STBY and disconnect all
test equipment.
5-17.
HP 5385A Channel 8 Input Amplifier
Adjustment Procedure
5-18.
To perform the sensitivity adjustments, pro-
ceed as follows:
a.
Cycle the power switch from STBY to ON. Press
FREQ B button on the front panel.
b.
Locate variable resistor A 1 R34 on the A 1 Mother-
board. Refer to the A
1
component locator in Section
VIII.
c.
Connect instruments as shown in
Figure 5-5.
5-3