Chapter 1 Performance Tests

Test 1: Time Interval (HP 53131A/132A Only)

Test 1: Time Interval (HP 53131A/132A Only)
This test verifies the Time Interval specifications between Channels 1 and 2.

Equipment

HP 8130A Pulse Generator (or equivalent)
HP 10503A BNC Cable

Counter Setup

1
Cycle the POWER key to preset the Counter.
2
Press Time & Period key until TI 1 TO 2 is displayed.
3
Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: .500V (Enter)
Trigger/Sensitivity, COMMON 1: ON
50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
4
Set CHANNEL 2 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: .500V (Enter)
Trigger/Sensitivity, SLOPE: NEG
50Ω/1MΩ, 50Ω
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Assembly-Level Service Guide
1-17