Chapter 1 Performance Tests
Test 3: Channels 1 and 2 Frequency Sensitivity (HP 53131A/132A Only)
100 kHz to 100 MHz Sensitivity for HP 53131A/132A
1
Connect the timebase output of the unit with the better timebase
to the timebase reference input of the other unit as shown in
Figure 1-7.
For this test, make sure you always use the unit (HP 53131A/132A or
HP 8663A) that contains the better 10 MHz timebase as the output source
as shown in Figure 1-7.
HP 53131A/132A
Counter
Channel 1
HP 53131A/132A
Counter
Channel 1
Figure 1-7. Channels 1 and 2 Frequency Sensitivity Test Setup
Assembly-Level Service Guide
HP 8663A
Synthesized
Signal Generator
50Ω
Feedthrough
Ref. In
(rear panel)
OR
HP 8663A
Synthesized
Signal Generator
50Ω
Feedthrough
10 MHz
Output
(rear panel)
Output
(rear panel)
N-to-BNC
Connector
Output
Input
(rear panel)
N-to-BNC
Connector
Output
1-23