Error Number 123 - Filtec FT-70c Service And Technical Reference Manual

Fill level inspector, for can inspection
Table of Contents

Advertisement

Industrial Dynamics Company, Ltd. © Form 40256-0600

Error Number 123

The X-ray Status Function is turned off. Turn the status back on to
activate x-ray generator.
Error Number 124
X-ray Source Disabled through external input connected to pin
TB4-5.
Error Number 125
X-ray Generator failure detected during warm-up cycle and the
generator will not activate. Check the LEDs on the generator.
• If green LED is off the generator is not receiving power from
system.
• Yellow Source Enabled LED should be on if unit is receiving
the signal to enable generation.
• If red LED is off and the green LED is on, the source voltage is
low.
• If red LED is on check the malfunction signal at pin TB4-5.
Error Number 126
X-ray generator is warming up for 3 minutes. This warm up
routine occurs if the generator is turned off for more than one
week. When the warm up s complete this error will automatically
clear.
Chapter 8: Diagnostic and Repair Procedures
8-13