Chapter 8: Diagnostic and Repair Procedures

Error Number 52

A container passed the trigger in less than 10ms. This can be
caused by trigger reflections or line jams.
To correct this problem:
Adjust the gain on the trigger amplifier.
Error Number 53
The Reject Watchdog detected a queue error. If this error occurs
repeatedly, call IDC technical support.
Error Number 54
This error displays the Reject Queue Count. If this error occurs
repeatedly, call IDC technical support.
Error Number 58
An abnormally large container was detected. The reject delay
encoder count exceeded the current encoder count. This condition
is generally caused by line jams. If this error occurs continuously,
the rejector may be installed to near the trigger.
Error Number 61
The reject delay position at the rejection queue output exceeded
the current encoder count. If this error occurs repeatedly, call IDC
technical support.
Error Number 66
A fill level reject was detected while the Fill Level Status was
turned off. Turn the Fill Level Status on to correct this condition.
Error Number 67
The number of containers in the Consecutive Diameter Missing
Cap Reject Alarm was exceeded.
This is a high priority error condition and will cause the Warning
Beacon to flash if Beacon Status is set to HIGH PRIORITY ONLY.