Replacing The Probe Module; Regular Care; Condensate Trap - TESTO 330-1 Instruction Manual

Flue gas analyzer
Table of Contents

Advertisement

E.2.2 Replacing the probe module
E.3 Regular care
E.3.1 Condensate trap
The fill level of the condensate trap can be read from the markings on the trap.
A warning message (
reaches 90%.
Emptying the condensate trap
The condensate consists of a weak mix of acids. Avoid contact with the skin. Make
sure that the condensate does not run over the housing.
Condensate in gas path.
Damage t t o m m easuring c c ells a a nd f f lue g g as p p ump.
Do not empty condensate trap while pump is operating.
Placeholder:
Kondensatfalle_oeffnen.tif
Kondensatfalle_leeren.tif
1 Press the key on the top of the probe handle and
remove the probe module.
2 Fit a new probe module and engage it in place.
red flashing light) is displayed if the level in the condensate trap
1 Hold the measuring instrument so that the
condensate outlet points up.
2 Open condensate outlet in condensate trap: Pull
out approx. 5mm or until it will not go any further
( ).
3 Let the condensate run out into a sink ( ).
4 Dab off drops at condensate outlet using a cloth.
5 Close the condensate outlet.
The condensate outlet must be fully closed (mark-
ing) otherwise incorrect measurements due to
inleaking air may result.
E. Operation
E.2 Probes / Sensors
19

Advertisement

Table of Contents
loading

This manual is also suitable for:

330-3330-2330-1 ll330-2 ll

Table of Contents