I. Care and maintenance
48
I.1 Cleaning the measuring instrument
I.

Care and maintenance

This chapter describes the steps and action required in order to keep the product func-
tioning properly.
See also Regular care , p. 19.
I.1

Cleaning the measuring instrument

If the housing of the instrument is dirty, clean it with a damp cloth. Do not use any
aggressive cleaning agents or solvents! Weak household cleaning agents and soap
water may be used.
I.2

Replacing measuring cells

A slot bridge (0192 1552) must be inserted in slots which do not have a measuring cell.
Used measuring cells must be disposed of as special waste!
The instrument must be switched off.
1 Place the measuring instrument on its front.
2 Remove the service cover: Take hold of it at the markings (arrows) using the index fin-
ger and thumb, press slightly, fold up and remove.
3 Pull hose connections from the fault measuring cell / bridge.
4 Remove the faulty measuring cell / bridge from the slot.
Placeholder:
Zellenwechsel.tif
Remove short-circuit brackets / additional boards
from the new measuring cells just prior to insertion.
Do not leave measuring cells for longer than 15
minutes without short-circuit brackets / additional
boards.
CO measuring cell for testo 330-1 / testo 330-1 LL:
Remove the shorting jumper
CO measuring cell for testo 330-2, -3 / testo 330-2LL:
Remove the shorting jumper
NO / NO
measuring cell:
low
Remove the auxiliary circuit board
.
.
.