Recalibrating/Adjusting Sensors; Modular Flue Gas Probe; Cleaning The Flue Gas Ducts; Replacing The Probe Module - TESTO 320 Instruction Manual

Flue gas analyzer
Hide thumbs Also See for 320:
Table of Contents

Advertisement

7.5.
7.6.
7.6.1.
7.6.2.
info@Testo-Direct.com
All manuals and user guides at all-guides.com
5. Install new sensor / new bridge in the slot.
6. Push the hose connections on to the sensor/bridge.
7. Refit and close the service cover.
After replacing an O2 sensor, wait for an equalisation
period of 60 min to elapse before using the instrument
again.
When retrofitting a sensor, the associated measurement
parameter and unit must be enabled, see Sensor settings,
page 40.

Recalibrating/adjusting sensors

See Recalibration/adjustment, page 41.

Modular flue gas probe

Cleaning the flue gas ducts

✓ Disconnect the flue gas probe from the measuring instrument.
1. Release the probe catch by pressing the key on the probe
handle and remove the probe module.
2. Blow compressed air through the flue gas ducts in probe
module and probe handle (see illustration). Do not use a brush!
3. Fit a new probe module on the handle and engage in place.

Replacing the probe module

✓ Disconnect the flue gas probe from the measuring instrument.
www.
Testo-Direct
7 Maintaining the product
61
1.888.610.7664
.com

Advertisement

Table of Contents
loading

Table of Contents