Specifications; Inputs; Control - Newport Pt32 Operator's Manual

Temperature, process and strain controllers, platinum series
Table of Contents

Advertisement

PLATINUM Series Controllers User's Guide

9. Specifications

9.1 Inputs

Input Types
Current Input
Voltage Input
Thermocouple Input
(ITS 90)
RTD Input (ITS 90)
Strain & Bridge Input
Excitation
Configuration
Polarity
Accuracy
Resolution
Input Impedances
Temperature
Stability
A/D Conversion
Reading Rate
Digital Filter
CMRR
Setpoint Adjustment
Warm Up to Rated
Accuracy

9.2 Control

Action
Reverse (heat), direct (cool), or heat / cool
Autotune
Operator initiated from front panel
Adaptive Tune
User selectable; fuzzy logic continuous PID tuning optimization
Control Modes
On/off or the following time/amplitude Proportional Control Modes:
selectable Manual or Auto PID, Proportional, Proportional with Integral,
Proportional with Derivative
Cycle Time
0.1–199 seconds
Thermocouple, RTD, Thermistor, Analog Voltage, Analog Current, Strain
4 to 20 mA, 0 to 24 mA Scalable
±50 mV, ±100 mV, ±1 V, ±10 Vdc Single Ended Scalable
±50 and ±100 mV differential and ratiometric inputs, scalable
K, J, T, E, R, S, B, C, N
100/500/1000 Ω Pt sensor, 2-, 3- or 4-wire; 0.00385 (100 Ω only), 0.00392
(100 Ω only), or 0.003916 (100 Ω only) curves
±50 mV, ±100 mV
Firmware selectable (no jumpers to set) to 5, 10, 12, and 24 Vdc @ 25 mA
(24V for AC units only)
Differential
Bipolar
Refer to Table 8
0.1°F/°C temperature; 10 µV process
Process Voltage/Strain: 10 MΩ for +/- 100 mV
Process Voltage: 1 MΩ for other voltage ranges
Process Current: 5 Ω
Thermocouple: 10 KΩ max
RTD: 0.04°C/°C
TC at 25°C (77°F): 0.05°C/°C (cold junction compensation)
Process/Strain: 50 ppm/°C
24 bit Sigma Delta
20 samples per second
Programmable from 0.05 seconds (filter = 1) to 6.4 seconds (filter = 128)
120 dB
4 Digit (-9999 to +9999 counts); 6 Digit (-99999 to +999999)
30 min
66
Specifications

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pt16Pt8dPt16dPt8ePt8pt

Table of Contents