Measurement Condition Settings; Measurement Condition Settings <Setting> Tab - Konica Minolta CA-410 Series Instruction Manual

Display color analyzer pc software ca-s40 ver. 1.3 probe+
Hide thumbs Also See for CA-410 Series:
Table of Contents

Advertisement

Measurement condition settings

Click [Settings] among the individual functions to bring up the following pop-up window for setting the measurement
conditions and optional items.
1. Measurement condition settings <Setting> tab
Measurement
mode
Calibration channel
information
1-1. Measurement mode settings [Mode tab - Mode Settings]
 Sync Mode
Measurement
synchronization
mode
NTSC
Mode for measuring NTSC displays
Mode for measuring PAL and SECAM
PAL
displays
Mode for measuring while
EXT
synchronizing with the display's vertical
synchronization signal input
Mode used when the display's vertical
scanning frequency is unknown or
UNIV
cannot be input; a mode that can
measure any display
When the display's vertical
synchronization signal is known, set
INT
that information so that it can be used
for measurement.
Mode that allows you to specify the
MANUAL
integration time
:
Select the mode for synchronized measurement. (NTSC/PAL/UNIV/INT/EXT/MANUAL)
Description of mode
Measurement time
(Sampling time)
33.3 ms
40.0 ms
(1 vertical scanning period)
x 2
100 ms
(1 vertical scanning period)
x 2
Settable range
4.0 to 4000.0 ms
(For FMA, 7.7 to 4000.0 ms)
54
<Settings> window
Vertical scanning frequency
59.94 Hz
50 Hz
0.50 to 240.00 Hz
(For FMA, 0.50 to 130.00 Hz)
Settable range
0.50 to 240.00 Hz
(For FMA, 0.50 to 130.00 Hz)
Use the [SYNC Assist] tab to
search for the optimum
frequency.
Vertical
synchronization
signal input
Not required
Not required
Required
Not required
Not required
Not required

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents