Target settings

The target values used for user calibration are written as the target in a calibration channel. If you want to set a separate
target for color difference assessment (this function is available with a data processor), you can set a target in a calibration
channel using [Set Target]. While any target data already written in that calibration channel is overwritten, the set correction
coefficients remain the same.
If you do not want to lose the target values used for user calibration, it is recommended that you set them in a channel
different from the one you used for user calibration.
• Click the [User Cal] button on the chromaticity measurement screen to open the pop-up window for user calibration.
• The calibration channel (CH00 to CH99) data is read before displaying the pop-up window.
• As calibration channel data, the ID, target probe, calibration probe, calibration mode, Lv,x,y, calibration date/time, and
comments are displayed.

1. Setting the target

Task selection tabs
Calibration mode
(1) Setting the target
(1)
Select the [Set Target] tab.
(2)
Select the calibration channel in which the target is set. (CH01 to CH99)
• Click over the calibration channel data or select from the channel selection drop-down menu.
(3)
Enter the target values in the [Measured Value] field or obtain the value by clicking the measurement button on the
right side. The values can be rewritten after measurement. Performing zero calibration by clicking the [0-CAL] button
is recommended before running a measurement.
(4)
Click [Apply] to complete setting the target. The change is not applied until you click [Apply].
(2) Setting the ID
To set the ID or enter a comment, follow the same procedure as for setting the ID and entering a comment for user
calibration. See 2-3 on P.63 for instructions on how to set the ID. (ID is set for a calibration channel.)
Calibration channel data
Measurement button
65
Selected channel
Setup