The Automatic Shutdown feature is designed to prolong the battery
life of the instrument when it is not in use. By default, the instrument
automatically shuts off after 1 minute has elapsed.
The Auto-Shutdown time may be increased or decreased by
pressing the CHANGE keys. The times available are 30 seconds
(0.5 minute), 1, 2, or 4 minutes. Once you have made your selection,
press NEXT.

Save Mode/Storing Test Results

The Save Mode menu is shown in Display # 22.
Display # 22
The OtoRead unit automatically stores only the most recent test result
for each ear (L/R). The unit has the capacity to store 250 individual
tests. It is recommended that you go to the Test Results Clear
screen (described on page 29) to clear any previous memory
locations after you first select this option. When the OtoRead unit has
a full memory (250 saved tests), it will not allow any further testing.
At this point either the results must be printed or they must be
cleared from memory. It is important to keep a record of the test
number for each patient if this feature is used.
Upon saving test number 45, the red "ERROR" light will come on. At
this point, five more tests can be stored. Press CONTINUE to
proceed with testing. When 250 tests are saved, select TEST from
the Main Menu, the display will ask if you want to clear all 250
results. (NOTE: You may print or download the saved results at this
time.) If you choose CLEAR, all 250 tests will be deleted and the test
numbers will start again at 1. The OtoRead unit will ask to verify your
decision to erase all stored tests.
OtoRead Screener
Changing Instrument Settings
Page 30