Error Codes - Planmeca Intra User Manual

Intraoral dental x-ray system
Hide thumbs Also See for Intra:
Table of Contents

Advertisement

ERROR CODES

18
ERROR CODES
ERROR CODE
E r. 0 0
E r. 1 0
E r. 1 1
E r. 1 2
E r. 1 3
E r. 2 9
E r. 3 0
E r. 3 1
E r. 3 3
E r. 3 4
E r. 3 6
E r. 3 7
E r. 3 8
E r. 5 0
E r. 5 1
E r. 5 2
E r. 5 7
E r. 6 0
E r. 6 1
E r. 7 1
E r. 8 1
E r. 8 3
50
Planmeca Intra X-ray Unit
The error code is displayed on the time display.
Press the SELECT key to clear the error from the display.
ERROR MESSAGE EXPLANATION
Exposure key was released too early during the exposure.
X-ray tube Anode voltage (kV) overshoot.
X-ray tube Anode voltage (kV) dropped suddenly.
X-ray tube cathode filament preheating voltages are not calibrated.
Filament preheating voltage calibration failed.
Membrane keyboard key short-circuited/pressed during the self test or
faulty display board.
kV value does not reach or it exceeds the given value (difference more than
5%).
X-ray tube Anode current (mA) missing, or not in specified limits.
X-ray tube Filament voltage (V) missing, or outside the range (too low or too
high).
X-ray tube Anode voltage (kV) missing, or below the specified limit.
Too long exposure.
kV feedback signal open circuit or short circuit.
mA feedback signal open circuit or short circuit.
Tube head temperature sensor short circuit.
Tube head temperature sensor open circuit.
Filament voltage feedback not in specified limits.
Exposure key pressed during self test.
± 15VDC voltage is out of limits.
Communication error between control panel and tube head CPU.
FLASH memory check-sum error (tube head CPU).
EEPROM memory defective (tube head CPU).
Config register error (tube head CPU).
User's Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents