CONTROL PANEL
Parameter adjustment keys
Ready indicator light
Exposure key
Exposure warning indicator light
14
Planmeca Intra X-ray Unit
down
up
NOTE Planmeca Intra with Dixi3 system:
Press the SELECT key briefly to select the parameter -
kV, mA, exposure time or density - to be changed. When
the parameter value is flashing on the display, the
parameter can be changed with the parameter adjustment
keys. The up key increases the value and the down key
decreases it.
After adjusting the kV or mA value or exposure time the
unit will return automatically to the time adjustment mode
after 5 seconds time.
The green ready indicator light will come on when the unit
is ready to take an exposure. The waiting time between
exposures is 15 times exposure time, but always at least
6 seconds.
In the programming mode the ready light will start to flash.
You can set the unit so that the Ready indicator light
will only come on when the Dimaxis program is ready
for the exposure, i.e. "Waiting for exposure" message
appears on the computer screen. To change the
settings of the unit contact your Planmeca technical
support.
When you take an exposure you must press and hold
down the exposure key for the duration of the exposure.
The yellow exposure warning light will come on when you
take an exposure. You will also hear an audible warning
sound during the exposure.
User's Manual