4

PLANMECA INTRA X-RAY UNIT - MAIN PARTS

4.1

General view of the X-ray

Generator box
Horizontal tube head stiffness adjustment screw
User's Manual
PLANMECA INTRA X-RAY UNIT - MAIN PARTS
Support arm
BW
READY
PRET
kV
mA
s
MODE
SELECT
Control panel
Adjust the stiffness of the tube head horizontal movement
by turning the adjustment screw on the support axle
manually or with a wrench tool (see next page).
The stiffness of the tube head horizontal movement has
been preadjusted at the factory, and can be changed by
the user, if necessary.
Turn the adjustment screw 0,5 - 1 rounds clockwise if you
want to tighten the tube head and 0,5 - 1 rounds
counterclockwise to loosen it.
NOTE Do not turn the adjustment screw too much
counterclockwise to avoid the screw to come loose.
Scale for cone angle
Tube head