Install the Degasser (G1379A)

Ensure the line power switch at the front of the degasser is off.
CAU TI O N
1 Place the degasser on top of the pump. Make sure that the two modules are
2 Connect one end of the remote cable (5061-3378) to the rear of the
3 The degasser accessory kit has a set of 4 solvent tubes (G1322-67300). Each
Degasser
OUTLET
A
B
C
D

Install the Solvent Cabinet

1 The solvent cabinet accessory kit has 4 bottle head assemblies
2 Connect a bottle head assembly to each of the degasser INLET ports. Use
3 Purge your system before first use (see "Get the System Ready for the First
1100 Series Nano Pump Service Manual
interlocked correctly.
degasser. Connect the other end of the cable to the remote port at the rear
of the pump.
tube is labeled A, B, C or D. Connect each solvent tube to its intended
OUTLET channel port on the degasser. Connect the other end of the solvent
tube to its intended port at the pump solvent selection valve. Follow the
guide below:
(G1311-60003).
the labels provided with each bottle head assembly to appropriately label
each bottle head assembly.
Injection" on page 17).
Pump Solvent Selection
Valve Port
to
A1 (left half, upper)
to
A2 (left half, lower)
to
B1 (right half, upper)
to
B2 (right half, lower)
Installing the Pump
1
15