Removing the Top Cover and Foam

Tools required
Preparations for this
procedure
1 Remove the front cover by pressing the clip fastener
on both sides of the cover.
(2
(1
1100 Series Nano Pump Service Manual
Screwdriver Pozidriv #1
Switch off nano pump at the main power switch.
Disconnect the solvent inlet tubes from the solvent selection valve or the
adapter at the active inlet valve.
Beware of leaking solvents due to hydrostatic flow.
Remove leak funnel with the waste tube.
Remove the solvent cabinet from the nano pump.
Clip fastener (1)
2 Unplug the power cord and move the lever towards
the power socket.
Repairing the Pump
Cover clip
131