Agilent Technologies 1100 Series Service Manual page 158

Nano pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

3
Repairing the Pump
7 Remove the pump drive. For this, disconnect first:
8 Place the new pump drive into the recess. Reconnect the connectors at the
9 Re-install the Z-panel and the damper. Reconnect the capillaries and
10 Re-install the mixing chamber to its holder.
11 Re-install the pump head. Reconnect the capillaries and the cable.
12 Re-install the flow sensor. Reconnect the capillaries and the cable.
13 Re-install the EMPV. Reconnect the capillaries, the waste tube and the
14 Replace the top foam section, optional interface board, front cover, and top
146
the mixing capillary.
the capillary going to the mixer.
the active inlet valve cable at the Z-panel.
the connector (P16, P11, P20) at the main board (see Figure 22 on page
135).
the connector (P8, P21 or P12, P14) at the main board.
main board.
connectors. (P16, P11, P20).
cable.
cover (see "Replacing the Top Cover and Foam" on page 154).
1100 Series Nano Pump Service Manual