Exchanging the Power Supply

Frequency
Tools required
Parts required
1 Remove the pump from the stack, remove the front cover, top cover and top
2 Remove the EMPV. For this disconnect first:
3 Remove the flow sensor. For this disconnect first:
4 Unclip the mixing chamber from its holder.
5 Remove the damper and the Z-panel. For this disconnect first:
6 Remove both pump assemblies. For this, disconnect first:
1100 Series Nano Pump Service Manual
If defective
• Screwdriver Pozidriv #1
• Wrench 1/4 inch
• Wrench 14 mm
• Wrench 7 mm
• Wrench 5 mm
Power supply 0950-2528
foam section (see "Removing the Top Cover and Foam" on page 131).
the capillary going to the flow sensor.
the capillary coming from the filter.
the waste tube.
the connector.
the capillary coming from the EMPV.
the capillary on the injection device (port 1).
the connector.
the mixing capillary.
the capillary going to the mixer.
the active inlet valve cable at the Z-panel.
the connector (P16, P11, P20) at the main board (see Figure 22 on page
135).
the capillary coming from the mixing chamber.
the connecting tube coming from the solvent selection valve.
the connector (P8, P21 and P12, P14) at the main board.
Repairing the Pump
147