Exchanging the Damper

Frequency
Tools required
Parts required
Do not totally remove the Z-panel when the damper is in its place. The heavy damper may
NO TE
fall down.
1 Remove the pump from the stack, remove the front cover, top cover and top
2 Disconnect both active inlet valve cables.
3 Remove the EMPV. For this disconnect first:
4 Remove the flow sensor. For this disconnect first:
5 Unclip the mixing chamber from its holder.
6 Remove the damper and the Z-panel. For this disconnect first:
7 Unclip the mixing chamber from its holder
8 Loosen the screws of the Z-panel, fold it forward and remove the damper.
9 Re-install the Z-panel and the new damper. Reconnect the capillaries and
1100 Series Nano Pump Service Manual
No pressure output or when leaking
Screwdriver Pozidriv #1
Wrench 1/4 inch
79835-60005 Damper
foam section (see "Removing the Top Cover and Foam" on page 131).
the capillary going to the flow sensor.
the capillary coming from the filter.
the waste tube.
the connector.
the capillary coming from the EMPV.
the capillary on the injection device (port 1).
the restriction capillary.
the capillary going to the mixer.
the connector (P16, P11, P20) at the main board (see Figure 22 on page
135)
connectors. (P16, P11, P20) (see Figure 22 on page 135)
Repairing the Pump
141
3